OPERATION
s NOTICE: The Neb-u-Lite LX2 Nebulizer Compressor is
designed for intermittent use only. Do not operate it
continuously for more than 30 minutes for a single use
without turning it off and allowing a cooling period of least
30 minutes.
1. Ensure that the ON/OFF switch is in the OFF position.
2. Plug the power cord into a properly grounded electrical
wall outlet.
3. Move the ON/OFF switch to the ON position to start the
compressor.
4. Place the nebulizer mouthpiece between teeth and
close mouth (or install the optional mask on face).
5. As aerosol starts flowing, inhale deeply and slowly
through mouthpiece (or optional mask).
WARNING: Never use a nebulizer with a clogged nozzle.
If the nozzle is clogged, the aerosol mist will be reduced,
altered or stopped, which will diminish or eliminate the
effectiveness of the treatment. If clogging occurs, replace
the nebulizer.
Info: If you need to interrupt treatment, move the ON/OFF
switch to the OFF position.
6. When the treatment is complete, move the ON/OFF
switch to the OFF position to turn off the compressor.
7. Unplug the compressor from power outlet.
8. Disconnect nebulizer from tubing.
9. Clean nebulizer as instructed in MAINTENANCE/
Cleaning the Nebulizer Cup.
JB0112-110-INS-LAB-RevD18
10