TROUBLESHOOTING
If your Neb-u-Lite LX2 Nebulizer Compressor fails to
function, consult the following Troubleshooting guide.
If the problem persists, consult your Graham-Field
authorized distributor.
Problem
Device doesn't
operate
Nebulization is
weak
JB0112-110-INS-LAB-RevD18
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to
Neb-u-Lite compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.
14