INSTRUCTIONS FOR USE
Setup:
a. The patient should be safely
positioned on the table and
secured accordingly.
b. Place the arm board post
into a clamp/socket that is
positioned appropriately on
the table siderail.
c. Once the Arm Support is
placed in the clamp and
the desired positioning is
achieved (based on the
position of the arm board
and clamp/socket as it
relates to the orientation of
the
patient),
secure the clamp/socket to
the table by tightening the
knob.
Patient Positioning
a. Once the patient is safely supported in the desired position, place the arm on
the padded Arm Support. Use the straps to secure the patient's arm to the
surface as needed.
b. Fine adjustments of the height of the Arm Support may be made by
supporting the patient's arm while loosening the post in the clamp/socket.
When properly positioned, the patient's arm will rest gently on the Arm Support
and the palm of the hand will "cup" the palm support. The arm is then
balanced in a passive, anatomically correct position. Be sure to re-tighten the
clamp/socket.
c. Further adjustments can be performed by repeating the above step in
addition to loosening the Pivoting Block Handle and repositioning the patient's
arm accordingly. Once the desired positioning has been achieved, be sure to
re-tighten all knobs and clamps/sockets.
Document Number: D-720515
Version: C
you
can
Page 12
Issue Date: 31 MAR 2020
Ref Blank Template: 80025118 Ver. E