Apex Cleco NeoTek 30 Serie Manual De Instrucciones página 12

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 27
1
EN
Malfunction
Warning in RUN SCREEN:
Tool Offline!
Different torque display in
the control and external
transducer
Tool cable does not fit in tool
handle
12
Possible causes
Tool has not been accepted
in the Tool Settings
A faulty cable will prevent
the tool from being recog-
nized by the control.
Torque calibration factor
(Torque Cal.) in Tool Set-
tings must be adjusted.
Customer's angle attach-
ment was added to tool.
Tool memory must be
updated.
30/50 series tools require a
30/50 series tool cable.
2019-02
TM12-1109 |
Remedy
 Check the Tool List menu to make
sure that the tool has been
accepted. If it has not been
accepted, accept it. The correct line
must be marked before the setting
can be accepted in the Tool Settings
menu.
 Repair or replace cable
 Adjust Torque Cal. (Tool Settings)
using the following formula:
 New calibration value = (external
value / tool value) x calibration value
used
 Reprogram tool memory
 Use the right cable for the tool being
used.
Apex Tool Group

Publicidad

Tabla de contenido
loading

Tabla de contenido