Ethylene Oxide (EO)
Preconditioning parameters
Temperature
Exposure
Concentration
(100% EO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
Steris System 1 / Steris System 1E
Note: Steris® System 1™ is not intended for use in the United States.
Do not sterilize the autoclavable coupler (1188-410-110) with Steris
System 1E.
1. Clean and prepare the coupler as recommended in this user guide.
2. Sterilize the coupler using Steris® System 1™ with Steris® 20
Sterilant, or System 1E™ with S40™ Sterilant. Follow the
manufacturer's instructions.
3. Allow the coupler to completely dry before reassembling it to a
scope or camera head. Otherwise, the lens will fog during use.
Sterrad
Caution: Couplers marked
sterilization. Steam sterilization is the preferred method.
1. Clean and prepare the coupler as recommended in this user guide.
2. Sterilize the coupler using the Sterrad™ NX or 100S System.
3. Allow the coupler to completely dry before reassembling it to a
scope or camera head. Otherwise, the lens will fog during use.
Storage
N/A

Technical Specifications

Transport and Storage
Conditions
Operating Conditions
8
55 ± 2°C (131 ± 4°F)
725 mg/L
131°F (55°C) ± 5°F
1 hour
70% RH (50 – 80%) ± 5%
12 hours
35 – 55°C (95° – 131°F)
are not suitable for Sterrad
AUTOCLAVE
Temperature: -10 – 50°C
Relative Humidity: 10 – 75%
Atmospheric Pressure: 700 – 1060 hPa
Temperature: 10 – 30°C
Relative Humidity: 30 – 75%
Atmospheric Pressure: 700 – 1060 hPa