ResMed produces a Data Management Guide for all compatible flow generators with
specific information on the data obtained from the R
S
OFTWARE
Software upgrades can be performed via the SmartMedia card for either R
the flow generator. The upgraded version of the software is shown on the label of the
SmartMedia card.
The SmartMedia card that is packed with the R
an upgrade for A
upgrades may be shipped separately.
If the flow generator has not yet been upgraded, the R
automatically conduct an upgrade when the SmartMedia card is first inserted. The initial
software upgrade ensures that the flow generator is compatible for use with the
R
L
ES
INK
patients. This will take approximately 20 minutes.
WARNING
!
Do not attempt to use the flow generator for therapy during the upgrade.
Note: If your AutoSet Spirit or S7 Elite has software version SX116-0207 or less, you will
need to follow these steps to upgrade:
Download any summary data from the flow generator to AutoScan or ResScan via cable.
•
Insert the SmartMedia card into the ResLink.
•
Connect the ResLink, power on the flow generator, and the upgrade will start
•
automatically.
Erase the summary data from the AutoSet Spirit using the LCD, AutoScan or ResScan.
•