Display
Alternate (blinking)
Alternate (blinking)
Error code
Sample mass error
The balance cannot store the sample for the counting mode
or for the percent mode because it is too light.
Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too light.
Storing and using it for counting will cause a counting error.
Add samples to reach the specified number and press the
PRINT key. Pressing the
samples will shift the balance to the counting mode. But, for
accurate counting, be sure to add samples.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and
calendar function works normally as long as the AC adapter
is connected to the balance. If this error appears frequently,
contact the local A&D dealer.
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Please check if the problem is the AC adapter supplied with
the balance.
ECL repeatability
With the self-check function, the standard deviation (SD) of
repeatability due to electronically controlled load (ECL)
exceeded 50 digits.
Please revise the installation environment of the balance.
This is displayed when repeatability is displayed by ECL.
This is displayed when the minimum weighing value
(reference value) by ECL is displayed.
Refer to the separate instruction manual "6-2 Self Check
Function / Automatic Setting Of Minimum Weighing Value by
ELC.
Full memory
The maximum number or stored weighing values has been
reached. In order to store more weighing values, it is
necessary to delete the data. Refer to the separate
instruction manual "11 Data Memory".
Full memory
The stored calibration and calibration test history has
reached 50 instances. If more is stored, the old history will
be deleted. Refer to the separate instruction manual "11
Data Memory"