CLEAN AND MAINTENANCE NOTICE
Before and after each using, please clean the following parts:
1
1. Remove the medical cup and mouthpiece or mask.
2. Discard all remaining medicine in the cup.
Open the medical cap and discard all remaining medicine.
Pour in some hot water and close the cover completely.
Press the power button and operate 1~2minutes to discard all remaining medicine completely.
Ensure medicine be cleaned completely after each using, otherwise, mesh can be blocked.
3. Clean the surface of main body of appliance with cotton swabs or a piece of fine cotton fabrics.
Please clean the electrode in the main body and medical cup, ensure normally electrical conduction.
Please do not use easily volatility liquid to clean the electrode, ex.benzene, thinner.
4. Clean the medical cup, mouthpiece and mask with clean water and store at a safety place after cleaning the
surface with ethanol.
Suggestion: Use a new medical cup for different illness or patients to avoid cross infection.
DISINFECTION
Medical cup can be disinfected by boiling in hot water (90°C):1 min, Validated number of cycles are 20. After 1 minute
boiling, shake off excessive water from the cup, and let dry on a clean, lint free(paper) towel. Main unit, mask, mouth
piece cannot be boiled.