Apex Mobi Mesh Manual De Instrucciones página 15

Ocultar thumbs Ver también para Mobi Mesh:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
12
Mobi Mesh
/ English
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The ME equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ME equipment can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the ME equipment as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
The ME system is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME system should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Vapor Expert (PY-001) Portable Nebulizer - Instruction Manual
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz d=1,2
P
0.12
0.38
1.2
3.8
12
The ME declaration – electromagnetic immunity
3V
3V/m
80 MHz to 800 MHz d=1,2
P
0.12
0.38
1.2
3.8
12
Electromagnetic environment - guidance
Portable and mobile RF communicationss equipment should be used no
closer to any part of the EQUIPMENT or SYSTEM including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter. Interference may occur in
the vicinity of equipment marked with the following symbol.
800 MHz to 2.5 GHz d=2,3
P
0.23
0.73
2.3
7.3
23