formers may damage the appliance.
MICROBIAL CONTAMINATION
In presence of pathologies with risks of infection and contamination, a
personal use of the accessories is suggested (always consult the physi-
cian).
CLEANING
1) Caution! Hot surface!
Before performing the cleaning of the device, let it cool down
since the transducer and the residual inhalation preparation be-
come hot during the use.
Always clean the appliance after each use in order not to let the
residual medicinal products solidify and damage the mechanism.
2) Before performing the cleaning, remove batteries or, in case of
operation through mains, disconnect the power supply plug.
Make sure the rubber membrane placed on the bottom of the de-
vice is closed to prevent any liquid penetration.
3) Remove the appliance cover and lift the cover of the atomisation
chamber to discharge possible residuals from the chamber itself.
Clean the atomisation chamber and the atomisation supplier as
follows:
- disconnect the atomisation chamber from the appliance by
pressing both hooks located on the bottom of the chamber itself
(See fig. 5). Leave the atomisation supplier connected to the at-
omisation chamber;
- pour 8 ml of ethyl alcohol in the atomisation chamber; let it act
for 10 minutes and then slightly shake the atomisation chamber.
Remove ethyl alcohol.
- pour 8 ml of water in the atomisation chamber; let it act for
10 minutes and then slightly shake the atomisation chamber.
Remove water.
- disconnect the atomisation supplier from the atomisation
chamber by unscrewing it counterclockwise and rinse it under
running water. Rinse also the atomisation chamber under run-
ning water.
- use a cotton fioc® moistened with ethyl alcohol to clean the two
golden screws present in the atomisation supplier. Rinse again
under running water and let dry. After the cleaning, screw the
atomisation supplier clockwise to the atomisation chamber until
the two hooks match. (Pay attention to the two golden screws,
which have to be positioned downwards).
Before hooking the atomisation chamber to the main unit, make
sure that it is perfectly dry. To assemble it again to the appliance,
place it on the main unit and press it slightly downwards until a
"click" is heard.
4) Clean the external part of the appliance using a soft cloth mois-
tened with water and neutral soap (Do not clean the main unit
placing it under running water and pay attention that no liquid
enters the appliance).
5) It is suggested to disinfect the accessories after each use to limit
the risk of infections. In this case, place mouthpiece, connection
part, and marks under running water and then in a 1/4 distilled wa-
ter and 3/4 alcohol solution for 10 minutes; then, rinse them and let
them fully dry before using them again.
Rinse them and let them fully dry before using them again.
6) Disinfect all components, also before using the device for the first
time.
7) The device, atomization chamber and all accessories must be
Mesh sonic nebulizer
completely dry before their use.
8) Store the device and all its accessories in a fresh and dry place in
the standard supplied case.
ATOMIZATION TIMES - FOR INFORMATION PURPOSES ONLY
• 4 cubic centimeter of physiological saline / water: 8 minutes
• 4 cubic centimeter of Cortinal®: 25 minutes
• 4 cubic centimeter of Fluibron®/Clenil®: 20 minutes 
TROUBLESHOOTING
• The supply of steam may be interrupted in the following cases:
- If during the treatment, the device has been tilted of more than 45°.
Keep the device in the correct position during inhalation. Read para-
graph INSTRUCTIONS FOR USE again.
- The drug present in the atomization chamber has almost run out.
The device is equipped with an automatic switchoff device that disa-
bles the device when medicine runs out.
- If batteries are flat.
When the orange warning light switches on, do not switch on the de-
vice, as by doing that damages may arise; replace batteries instead.
- The liquid to be inhaled shows some bubbles.
Remove the bubbles present in the atomization chamber and then
press the ON/OFF key.
• The device does not work correctly after a long period of standstill:
- Check that in the atomization chamber there are not any residuals of
drug of the last treatment.
Clean or dry the atomization chamber with a soft and dry cloth.
Check that in the atomization chamber and in the accessories there
are not any residuals of drug of the last treatment. Clean and dry the
components with a soft and dry cloth.

TECHNICAL SPECIFICATIONS

• Power supply: 4 alkaline batteries 1,5v AA
• Operating conditions:
Temperature: min. 5°C; max. 40°C
Relative humidity: less than 80%, in absence of conden sate
• Storing conditions:
Temperature: min. 10°C; max. 40°C
Relative humidity: less than 80%, in absence ofcondensate
• ultrasonic frequency: 100 KHz
• Nebulization capacity: about 0.25 ml/minute
• Medium size of the particles: 4.8 μm
• Maximum volume of the inhaling product: max. 8 ml = 8 cc
• Size: 7,5 x 7,5 x 14,4 cm
• Weight: 300 g
• Battery life: about 100 minutes
Device of BF type
Device of IIa class
Caution: check the instructions for use
DISPOSAL PROCEDURE (DIR. 2002/96/EC - WEEE)
The symbol on the bottom of the device indicates the separated col-
lection of electric and electronic equipment. At the end of life of the
device, do not dispose it as mixed solid municipal waste, but dispose it
referring to a specific collection centre located in your area or returning
it to the distributor, when buying a new device of the same type to be
used with the same functions. This procedure of separated collection of