Tabla de contenido

Publicidad

Quick Start

1.
Your recorder has been preset to operate using the most popular settings. To change the recording
range, input range or recording time, please reference Dip Switch Set-Up in this manual.
Recording Time: 7 day
Range: 0 to 100
Input Range Channel 1(Red Pen): 0 to 5V
Input Range Channel 2(Blue Pen): 0 to 5V
2.
Insert 9V battery and plug in AC adapter. (see figure 3) Plug in transducers using uncompensated
jacks. (see figure 4) NOTE: Be sure to observe proper polarity when connecting wires from transmitter.
(see figure 3)
3.
Remove the protective pen caps.
4.
The instrument will move pens to the appropriate readings.
5.
Install the chart that matches the Dip Switch settings you have selected.
a.
Press the PEN HOME key to move the pen to the outside of the chart. The pen is automatically
raised off the chart.
b.
Remove the old chart, place the new chart on the Chart Hub – being certain that the edge of the
chart slides under the Chart Guide Clips located at the outside of the chart.
6.
Set the appropriate time. There are two ways to adjust the chart and set the appropriate time:
a.
Set the chart time manually by inserting a coin into the groove in the chart hub and turning
clockwise until the correct hour (and day if applicable) on the chart is referenced to the timing
arrow (just to the right of the pen tip on the dial).
b.
(This feature should be used for fine adjustments only) To adjust the chart time, press and hold
ADJUST-UP and ADJUST-DOWN buttons located on the back of the unit next to the dip switch
(see figure 2). The green led will blink rapidly for about five seconds, then led will remain solid
green. While in this state ADJUST-UP button will move the chart forward (clockwise) and
ADJUST-DOWN button will move the chart backward (counter clockwise). Rotate the chart until
the correct hour (and day if applicable) on the chart is referenced to the timing arrow. Once you
have set your chart, press the PEN HOME button to exit Chart Adjust Mode. The unit will take
one minute to exit Chart Adjust Mode once PEN HOME is pressed.
7.
Press PEN HOME to move the pens back onto the chart.
8.
Place the ET8 recorder on a flat vibration-free surface. Be sure it is in a vertical position and level. For
best performance and longevity, the location should be a clean environment, free from dust and
corrosive fumes. Do not exceed temperature specifications.
Wall Mount: Keyhole slots are provided on the ET8 recorder for wall mounting.
Button Operations
Figure 2
Rev. 10/07
PEN HOME
PEN
(PWR)
HOME
LED
ON/ OFF
ALARM
ON/OFF
ALARM
Continued on next page

Publicidad

Tabla de contenido
loading

Tabla de contenido