Technical Specifications
Maximum heater plate
temperature
Temperature cut-out
Maximum output humidity
Maximum gas temperature
Dimensions (H x W x D)
Water chamber capacity to
maximum fill line
Weight (Humidifier and unfilled
water
chamber)
Power supply. Input range for
Tango™ flow generator with
humidifier
Instantaneous peak power
consumption
Maximum heater element power
Operating temperature
Operating humidity
Storage and transport temperature
Storage and transport humidity
Atmospheric pressure range
IEC 60601-1 Classification
Electromagnetic Compatibility
* Actual average power consumption will vary, depending on factors such as
temperature setting, set pressure, the use of accessories, height above sea level and
ambient temperature.
The manufacturer reserves the right to change these specifications without notice.
10
85ºC (185ºF)
93ºC (199ºF)
95% relative humidity (control dial set to maximum
setting, flow rate 50 L/min, ambient temperature of
20°C (68°F), ambient relative humidity of 40%)
41°C (106°F) at 35°C (95°F) ambient temperature.
87 mm x 189 mm x 409 mm (3.4" x 7 .5" x 16")
375 mL
920 g (2 lb)
100-120V/220-240V, 50/60Hz; <90 VA (80 W)
(maximum power consumption) *
<340 VA
120 VA (100 W)
+5°C to +40°C (+41°F to +104°F)
10–95% non-condensing
-20°C to +60°C (-4°F to +140°F)
10–95% non-condensing
1060 hPa (sea level) to 680 hPa 2591 m (8500 ft)
Class II (double insulation), Type CF
The C-Series Heated Humidifier is rated to IPX1 (The
power cord is excluded from this classification)
Refer to your flow generator user guide for details.