DiST®3 EC Tester
2
Recommended measurements
The coolant should be changed when the following values are exceeded:
Application
Plasma
TIG
MIG MAG
Tab. 1
3
Calibrating the EC Tester
To ensure that the EC Tester measures reliably and accurately, it should be calibrated weekly at minimum.
Rinse the surface of the EC Tester used to take measurements with water and dry it before each calibration.
Use new, clean calibration solution for each calibration.
NOTICE
• You must always use calibration solution 1413 µS/cm when calibrating.
• It is not possible to calibrate the device using coolant.
 Title page figure
1 Fill a clean, dry container with the supplied calibration solution (fill level min. 5 cm).
2 Press the On/Off button (3) to turn on the EC Tester.
3 Remove the EC Tester's protective cap.
4 Press the CAL button (2) until the "CAL" icon on the display (1) flashes and "µS/cm 1413 USE" is
displayed.
5 Submerge the EC Tester in the calibration solution up to the recommended immersion depth (4).
6 Wait until "REC" appears on the display (1).
7 Wait until "Stor" appears on the display (1).
The calibration is saved and the EC Tester returns to normal measurement mode.
BEI.0330.0 • 2019-03-11
Recommended measurements
0–30 µS/cm (note the manufacturer's specifications)
0–250 µS/cm
0–350µS/cm
Recommended measurements by application
2 Recommended measurements
EN - 3