Electromagnetic Compatibility (Emc); Emitted Interference Emc - Photonic ATO ML 600B Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Electromagnetic Compatibility (EMC)

Electromagnetic fields may interfere with the light.
Likewise, the light may affect other medical units
and impair their function. Therefore, observe the
limit values specified here!
Attention: These limit values only apply to the
lights in combination with the approved support
arm systems and accessories (see page 2). The
use of other or additional components is strictly
prohibited!

Emitted Interference EMC

Please note that the light may only be
operated in an electromagnetic environment as
specified below (see table)!
Phenomenon
Emitted
interference
3
Distortion
4
Flicker
1
Environment in home health care settings (including
professional health care facilities).
2
Power guided and radiated interference emissions
3
Harmonic distortion
4
Voltage fluctuations and flicker
Standard
section
CISPR 15
2
IEC 61000-3-2
IEC 61000-3-3
Test level
40
1
Group 1,
Class B
Class A
compliant

Publicidad