adjustment automatically corrects the deviation. At
the same time, a data record is printed out.
Important Note:
If the tolerance limit is too high, the analyzer will
not be adjusted. In this case, please inform your
local Sartorius office, dealer or service center.
§ Open the sample chamber by pressing the k
key
> Allow the moisture analyzer to cool down.
§ Snap the disk holder into the adjustment disk and
remove the disk. Caution: Equipment is extremely
hot after drying!
Do not touch to avoid severe burns!
> Allow the adjustment disk to cool on a clean, fire-
proof pad.

1-Point Adjustment

You should select the 1-point adjustment for samples
that are usually measured at one temperature.
When adjusting the temperature using the 1-point
adjustment, you can individually pre-select the
temperature. Use the number keypad to enter the
temperature. The factory setting is 105°C.
The routine corresponds to the 2-point adjustment.