Beurer IH50 Instrucciones Para El Uso página 30

Ocultar thumbs Ver también para IH50:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
Important
Before you clean the device, always
switch it off, unplug it and let it cool off.
Do not use abrasive detergents and never
immerse the device in water.
Important
• Ensure that no water penetrates inside
the device!
• Do not put the device or accessories in a
dishwasher.
• Do not touch the device with wet hands
while it is plugged in; do not allow any
water to be sprayed onto the device.
The device must be operated only
when it is completely dry.
• Do not spray any liquids into the venti-
lation slots.
Any liquid that penetrates into the de-
vice can damage the electrical parts or
other nebulizer components and impair
the function of the device.
Disinfection
• First clean the nebulizer and the acces-
sories as described in the Cleaning sec-
tion. You can then continue with disin-
fection.
• Pour 8 ml ethyl alcohol into the medi-
cine container (2). Close the container.
Leave the alcohol in the container for at
least 10 minutes.
• For better disinfection, shake the device
gently from time to time.
• Afterwards, pour the ethyl alcohol out of
the container.
• Repeat this process, using water this
time.
Disassembly
• To continue with cleaning its individual
components, disassemble the nebulizer
(2, 3):
• Remove the medicine container (2) and
mesh nebulizer (3) by pressing both
brackets inwards and drawing the de-
vice out.
• Remove the mesh nebulizer (3) from
the medicine container (2) by turning it
anticlockwise and drawing it out.
• The nebulizer is later reassembled in
the reverse order.
30

Publicidad

Tabla de contenido
loading

Tabla de contenido