ELECTROMAGNETIC COMPATIBILITY

Guidance and manufacture's declaration - electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Emmission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge (ESD)
± 8 kV contact
IEC 61000-4-2
± 15 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
line(s) to line(s):
IEC 61000-4-5
±1 kV. line(s) to earth:
±2 kV. 100 kHz repetition frequency
Voltage dips, short
5% UT (>95% dip in UT)
interruptions and voltage
for 0.5 cycle 40% UT (60% dip in UT)
variations
for 5 cycle 70% UT (30% dip in UT)
on power supply
for 25 cycle <5% UT (>95% dip in UT)
input lines
for 5s
IEC 61000-4-11
Power frequency (50/60 Hz)
3 A/m
magnetic field
50Hz/60Hz
IEC 61000-4-8
For all ME EQUIPMENT and ME SYSTEMS
Compliance
Electromagnetic Environment Guidance
Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
This device is suitable for use in all establishments, including domestic establishments and those directly connected to
Class B
the public low-voltage power supply network that supplies buildings used for domestic purposes.
Class A
Complies
For all ME EQUIPMENT and ME SYSTEMS
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
5% UT (>95% dip in UT)
for 0.5 cycle 40% UT (60% dip in UT)
for 5 cycle 70% UT (30% dip in UT)
for 25 cycle <5% UT (>95% dip in UT)
for 5s
3 A/m
50Hz/60Hz
Table 1
Table 2
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of this device requires continued operation during
power mains interruptions, it is recommended that this device be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
ENGLISH • 17