Caution!
•For patients with a weakened immune
system or a special medical history, daily
disinfecting or sterilization of the mask
components may be necessary. In this case,
please consult your doctor.
•Please take great care when cleaning and
disinfecting the contact surfaces between the
mask frame and the retaining ring. Residues
on these surfaces can reduce the success of
the therapy and the functioning of the
integrated exhaled air system (vented
variants).
Weekly
Disassemble the mask as described in section
3.12 "Disassembling/assembling the mask"
and subject the parts of the mask to a hygiene
treatment (see "4.2. Cleaning, disinfection,
sterilization" on page 20).
4.2 Cleaning, disinfection,
sterilization
Permissible procedures
Procedure
Parts
Mask cushion
Mask cushion GEL
Forehead cushion
Forehead cushion
GEL
Port cap
Mask frame
Coarse adjustment
component
Fine adjustment
component
Elbow
Anti-asphyxia valve
Rotating sleeve
Retaining ring
Headgear clip
Headgear
HEADstrap
20
Procedure
Parts
Headgear
JOYCEstrap
Quick release cord
Caution!
•Always clean the components of the mask
thoroughly before disinfecting or sterilizing
them as indicated in the above table.
•The emergency exhaled air valve and gel-
filled mask components are highly sensitive
components. Always handle them with care
to prevent damage or deformation.
Description of procedures
Rinse cycle at 65°C. Then carefully rinse
Dishwasher
all parts with clean water.
Hand wash in warm water. Use a
Hand wash
mild detergent. Then rinse all parts in
clean water and allow to air-dry.
Clean mechanically, e.g. with a brush,
in dilute solution. After disinfection rinse
Disinfection*
all parts thoroughly with distilled water
and dry thoroughly.
Sterilization
Steam sterilization in devices to comply
(alternative to
with EN 285. Temperature 134 °C,
disinfection)
minimum hold time 3 minutes.
*We recommend the disinfectants
GIGASEPT FF, CIDEX OPA and
ANIOSYME DD1. Evidence of tolerance to
these disinfectants has been provided. It is
essential to follow the manufacturer's
instructions.
Notes on the hygiene treatment and on the
hygiene treatment for a change of patient can
be found in a brochure from our information
portal at www.weinmann.de. If you do not
have Internet access, contact Weinmann.
Note
Wear suitable gloves (e.g. disposable
gloves) during the hygiene treatment.
Wash the HEADstrap headgear before us-
ing for the first time, as the colour might run.