Chicco TIRALATTE ELETTRICO Manual De Instrucciones página 25

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Proximity fields from RF
wireless communications
equipment
RATED power frequency
magnetic fields.
a) The equipment is suitable for use in Home Health Environment and Professional Health Care Environment limited to patient
rooms, breastfeeding and nursery facilities in hospital or clinics. The more restrictive IMMUNITY acceptance limits have been
considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance between the Electric Breast Pump mod. 06836
sources of power frequency magnetic field of at least 15 cm.
d) Reference should be done to clause 8.10 of EN 60601-1-2: 2015 and corresponding Table 9.
Guidance and manufacturer's declaration – Electromagnetic immunity – Input a.c. port
Phenomenon
Electrical fast transients /
Bursts.
Surges
Line-to-line
Surges
Line-to-ground
Conducted disturbances
induced by RF fields.
Voltage dips.
Voltage interruptions
)
c
a) r.m.s., before modulation is applied.
b) The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553
MHz to 13,567 MHz; 26,957 MHz to 27,283MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15
MHz and 80 MHz are 1,8 MHz to 2,0MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15
MHz, 14 MHz to 14,2MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz lo 21 ,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7
MHz and 50,0 MHz to 54,0 MHz.
c) ME EQUIPMENT with a d.c. power input intended for use with a.c.-to-d.c. converter is tested using the converter that
meets the specifications of the MANUFACTURER of the ME EQUIPMENT. The IMMUNITY TEST LEVEL is applied to the
a.c. power input of the converter.
IEC 61000-4-3
IEC 61000-4-8
Basic EMC standard or
test method
IEC 60601-4-4
IEC 60601-4-5
IEC 60601-4-5
IEC 60601-4-6
IEC 60601-4-11
IEC 60601-4-11
25
COMPLY
b
)
NOTE: Further detailed information on separation distances
between portable and mobile RF communications equipment
(transmitters) and the Electric Breast Pump mod. 06836
00009199000000 can be requested to Artsana S.p.A. throu-
gh the contact indicated in the present manual. However, it
is advisable to keep the Electric Breast Pump mod. 06836
00009199000000 at adequate distance of, at least, 1 m from mo-
bile phones or other RF communications transmitters in order to
reduce possible interferences.
30 A/m
)
c
50 Hz or 60 Hz
Livelli prova di immunità
Professional healthcare
facility environment
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ±1 kV
Class II equipment or supplied by Class II AC/DC adapter. Test
not applicable
± 0.5 kV, ±1 kV; ±2 kV
3V
)
a
0,15 MHz - 80 MHz
6 V a) in ISM and amateur radio bands between 0,15 MHz
and 80 MHz
b
)
80% AM at 1 kHz
0 % UT; 0,5 cycle
At 0°,45°,90°,135°,180°, 225°, 270° and 315°.
0% UT; 1 cycle
And
70% UT; 25/30 cycles
Single phase: at 0°
0% UT ; 250/300 cycles.
00009199000000 and
HOME HEALTHCARE
ENVIRONMENT
a
)

Publicidad

Tabla de contenido
loading

Tabla de contenido