If a container is used during weighing:
Place the empty container on the weighing surface until the stability indicator
then press the button marked
When the container is removed from the weighing platform, the digital display will show a
negative value. Place the object(s) to be weighed in the container, then place the
container on the weighing surface. When the stability indicator is shown
weight will then be displayed (without the container).
Press the button marked
mode.
Calibration
o Calibration may be required after the device is set up, or when it has been
transported a long distance.
o It is necessary because the the weight of a mass in one place is not always equal to
the weight in another place. Time and use may also result in decalibration.
o Before entering calibration mode, turn the device on. Turn it off after about 1
minute.
o Next, simultaneously press and hold the buttons marked
device on again. Release the buttons. After a test phase, the device will enter
calibration mode and the digital display will read "CAL". After 2 seconds the digital
display will read "0.000" or the value previously used for calibration.
o The displayed value cannot be changed. Press buttons
calibration weight appears.
o Next, place the displayed weight of calibration on the weighing platform and wait
until the stability indicator
o Press and hold the button marked
When the weight used for calibration is displayed, the process is completed.
o If the value exceeds ± 10% of the weight entered, the digital display will read "Err c".
Repeat the calibration process.
- 26 -
to set the value to "0.000".
, to reset the display to "0.000" and cancel this weighing
appears on the digital display.
until the calibration weight is displayed.
is shown,
, the net
and
then turn the
,
or
until the desired