 To Launch the cycle:
Press the
touch control, all the panel indicators light up.
Each of these indicators will go out at the appropriate time during the heating cycle.
N.B.:
.To move the program manually from one stage to the next, use the
. To cancel an active cycle, press the
CALIBRATION
All UGIN'DENTAIRE furnaces have been subjected to an IPTS 68 (International Practical Temperature
Scale of 1968) temperature test using a temperature calibrator.
However, should you not find the initial setting satisfactory; you can test the temperature using a pure
silver wire. Pure silver melts at 960°C.

TEST PROCEDURE

-Place the silver wire on a ceramic support in the middle of the muffle.
- Raise the temperature by stages. starting at 940°C with a 5 min hold, then 950°C/5 min, then 960°C/
5 min and, If need be, 970°C/min, until the test silver melts. (Open the furnace door briefly after each
stage to check on the silver wire).
Let us assume, for example, that the wire melted at 950°C. In that case, the calibration should be
modified as follows:
Press
and then
The figure 5 appears on the main display and the "Calib" indicator lights up.
Using the
key, raise the position number to 6.
Press
again.
N.B:
.Each position corresponds to a temperature change of 10° C relative to the starting point.
Hence, it is normal to observe that same deviation from ambient temperature when the furnace is
cold.
touch control.
26
PROGRAMIX 25/50/100