Descargar Imprimir esta página

König Electronic HC-EARTHERM60 Manual De Uso página 2

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Please read the instruction carefully before using
The ear thermometer is a device capable of achieving infra-red
temperature measurement when placed in the auditory canal of
a subject. It is a safe mean of measuring the human core body
temperature through the ear.
Specification:
The standard we adopted is EN12470-5:2003 Clinical thermometers
-- Part 5: Performance of infra-red ear thermometers (with maximum
device).
This device complies with the requirements of IEC 60601-1:1998, and
IEC 60601-1-2:2001.
Range of displayed temperature : 34.0°C ~44.0°C (93.2°F ~111.2°F)
Display L (L) when the temperature is under 34.0°C (93.2°F)
Display H (H) when the temperature is over 44.0°C(111.2°F)
Operating environment:
Temperature: 16°C ~35°C
(60.8°F ~95.0°F )
Humidity: ≤ 80%RH
Storage and transport conditions:
Temperature: -25°C to
55°C (-13°F to 131°F)
Humidity: ≤ 85%RH
Temperature scale: 0.1°C (0.1°F )
Accuracy: ± 0.2°C (from
35.5°C to 42.0°C)
± 0.4°F (from
95.9°F to 107.6°F)
Display: liquid crystal display, 3½ digits
Temperature value: display the maximum temperature in measuring
process
Temperature unit: centigrade or Fahrenheit
Memory: Last 10 measurements
Low voltage warning: the LCD display shows:
is heard
ENGLISH
Probe lens
On/recall button
Battery cover
2
Scan button
LCD
and then a beep sound

Publicidad