INSTRUCCIONES GENERALES DE PUESTA EN FUNCIONAMIENTO
HOSPIVAC CABINE con PROCOM 2 o CYCLIC 2020
SECCIÓN 14.
The vacuum units with PROCOM2 or CYCLIC 2020 automate needs special precautions regarding EMC.
The vacuum units with PROCOM2 or CYCLIC 2020 PROCOM2 automate has to be installed and put into service
according to the EMC requirements joined in the following tables
Guidance and manufacturer's declaration – electromagnetic emissions – for all ME equipment and ME systems
The vacuum units with PROCOM2 or CYCLIC 2020 automate is intended for use in the electromagnetic environment specified
below. The customer or the user of the vacuum units with PROCOM2 or CYCLIC 2020 automate should assure that it is used
in such an electromagnetic environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity - for all ME equipment and ME systems
The vacuum units with PROCOM2 or CYCLIC 2020 automate is intended for use in the electromagnetic environment specified
below. The customer or the user of the vacuum units with PROCOM2 or CYCLIC 2020 automate should assure that it is used
in such an electromagnetic environment.
Immunity test
IEC 60601 test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
IEC 61000-4-4
±1 kV for input/output
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
<5 % UT (>95 % dip in
Voltage dips, short
UT) for 0,5 cycle
interruptions and
40 % UT (60 % dip in
voltage variations on
UT) for 5 cycles
power supply input
70 % UT (30 % dip in
lines
UT) for 25 cycles
IEC 61000-4-11
<5 % UT (>95 % dip in
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Manual n° 517391 versión 5
ELECTROMAGNETIC COMPATIBILITY
Compliance
Group 1
Class B
Class A
Complies
Compliance level
For 230 Vac / 50Hz
±6 kV contact
±6 kV contact
±8 kV air
For 230 Vac / 50Hz
2 kV for power supply
lines
±1 kV for input/output
lines
For 230 Vac / 50Hz
±1 kV mode differential
±2 kV mode common
Pour 230 Vac / 50 Hz
0 Vac during 10 ms
92 Vac during 100 ms
161 Vac during 1 s
0 Vac during 5 s
UT) for 5 s
3 A/m
3 A/m for 50Hz
Electromagnetic environment – guidance
The vacuum units with PROCOM2 or CYCLIC 2020 automate
uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The vacuum units with PROCOM2 or CYCLIC 2020 automate is
suitable for use in all establishments, including domestic
establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for
domestic purposes
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
±8 kV air
Mains power quality should be that of a typical
lines
commercial or hospital environment.
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the vacuum units with PROCOM2 or CYCLIC
2020 automate requires continued operation
during power mains interruptions, it is
recommended that the vacuum units with
PROCOM2 or CYCLIC 2020 automate be
powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
No components sensible in product to magnetic
field.