CONNECTION AND ASSEMBLY
INSTRUCTIONS
1
2
1
Glass part of the electrode
2
Metal part of the electrode
● During assembly and disassembly of the high
frequency electrode, handle the area marked with a
dashed line, keeping a distance of 30 mm from the
metal part of the electrode.
● Connect the power cable to the rear panel of the
main body. A blue LED will automatically light up
around the main unit and the display screen and
parameter control will be activated.
● Connect the cable of the high frequency handpiece
to the corresponding output on the rear panel.
POWER AND CONFIGURATION
● After connecting the power supply cable and the
blue LED has lit up around the main unit, the
language selection menu will be activated on
the display screen: select the desired language
(Spanish or English).
● Press [POWER] on the control panel to turn off the
device. The selected language will be memorised.
● Press the [POWER] button on the control panel. On
the display screen and parameter control:
- The timer will show 15 minutes.
- The power indicator will show 1.
- The pause indicator
screen.
● Set the timer using the
the operating time of the machine, in a range of
between 1 and 30 minutes.
● Set the power using the
the power of the machine, in a range of between 1
and 9.
● Use the
button or press on the display screen
to start or stop the device.
- When in operation, an animation will be shown on
the display screen.
- When it is paused or on standby, the animation
will stop and the
display screen.
● To change the previously memorised language when
starting the device for the first time, disconnect
the power cable and wait for 10 seconds. After
reconnecting, the language selection menu will
appear again.
4
will appear on the display
or
buttons to choose
or
buttons to choose
indicator will appear on the
OPERATION
● Connect the device and leave it in standby mode
with the
indicator lit.
● Direct electrotherapy:
- Select the appropriate electrode for the
treatment, set the timer and decrease power to
the minimum.
- Hold the high-frequency handpiece with your
hand and press
An animation will automatically be shown on the
display screen and the device will start up.
- Then bring the electrode closer to the area of the
skin to be treated until it meets the skin and set
the power
gradually.
- At the end of the treatment, reduce the power
and press the button
screen to stop the device.
- When it is paused or on standby, the animation
will stop and the
indicator will appear on the
display screen.
- Then you can remove the electrode from the
patient's face.
● Indirect electrotherapy:
- The procedure is the same as above, except that
the patient should hold the straight electrode in
their hand from the beginning of the treatment.
- The operator's electrode should not come into
contact with the patient's electrode.
- During the treatment, the operator should keep
one hand on the patient's face.
Bear in mind:
● Humming sounds indicate that the high frequency
electrode is working normally.
● If using direct electrotherapy, when the electrode
is less than 2.5 cm away from the skin, a spark will
occur. This is why you should set the power to the
minimum and increase gradually. If the spark still
occurs, you can use a dry cotton cloth to cover the
face.
● [POWER] should be off when placing and removing
the high frequency electrode.
CLEANING AND MAINTENANCE
● Before cleaning the device, unplug it from the
mains.
● Clean the device with a soft, dry cloth. Do not use
liquids to clean it.
● After each use, the various components and
accessories must be sterilised and left to dry.
● For maintenance or repairs of any part of the
product, please contact your distributor or
specialised technician.
or tap on the display screen.
or tap on the display