Specifications
1000 measurement results with
Memory
respective date and time
Dimensions
98 (L) x 55 (W) x 15.5 (H) mm
Power Source
One 3.7V Li-Ion rechargeable battery
Weight
63.2g (without battery)
External output
3G/4G Mobile network
Auto electrode insertion detection
Auto sample loading detection
Auto reaction time count-down
Features
Auto switch-off after 3 minutes without
action
Temperature warning
8°C to 45°C (46.4°F to 113°F), below 85%
Operating Condition
R.H. (non-condensing)
Storage/
-20°C to 60°C (-4°F to 140°F), below
Transportation
95% R.H
Condition
Blood Glucose Test: mg/dL or mmol/L
ß-ketone Test: mmol/L
Measurement Units
Total cholesterol Test: mg/dL
Uric acid Test: mg/dL
Measurement Range
Hematocrit range
Test Sample
Test Result
This device has been tested to meet the electrical and safety
requirements of: IEC/EN 61010-1, IEC/EN 61010-2-101, EN
61326-, IEC/EN 61326-2-6, EN 301 489-17, EN 300 328.
EN-23
Blood Glucose Test: 10 ~ 600 mg/dL
(0.55 ~ 33.3 mmol/L)
ß-ketone Test: 0.1 ~ 8.0 mmol/L
Total cholesterol Test: 100 ~ 400 mg/dL
(2.6 ~ 10.4 mmol/L)
Uric acid Test: 3 ~ 20 mg/dL (0.179 ~
1.190 mmol/L)
Blood Glucose Test: 0 ~ 70%
ß-ketone Test: 10 ~ 70%
Uric Acid Test: 20 ~ 60%
Total Cholesterol Test: 20 ~ 60%
Blood Glucose Test: Capillary/ Venous/
Neonatal/ Arterial
ß-ketone Test: Capillary/ Venous
Uric Acid Test: Capillary/ Venous
Total Cholesterol Test: Capillary
Measurements are reported as plasma
equivalents