Table 2 Guidance and manufacturer's declaration – electromagnetic immunity – for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The BPW-9154 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BPW-9154 should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for
transient/burst
power supply lines
IEC 61000-4-4
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage dips, short
<5% UT
interruptions and
(>95% dip in UT )
voltage variations on
for 0.5 cycle
power supply input
40% UT
lines
(60% dip in UT )
IEC 61000-4-11
for 5 cycles
70% UT
(30% dip in UT )
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency 3A/m
3A/m
(50Hz) magnetic
field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 test level
Compliance level
±6 kV contact
±8 kV air
N/A
N/A
N/A
N/A
N/A
N/A
3A/m
13
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
GB