CONNECTION AND SET UP INSTRUCTIONS
WARNING
• Place the humidifier system on a hard and
flat surface only.
• Do not place the 9200 where it is accessible
to children or pets.
• Do not place the 9200 on top of the CPAP device/ventilator.
• The humidifier system must always be at a lower level than the mask and
the CPAP device/ventilator to ensure that condensation in the tubing flows
back to the humidifier chamber.
• If the humidifier system is placed on the floor, do not position the system
where it may be stepped on when getting out of bed.
• To ensure that the system airflow does not splash water from the humidifier
chamber into the delivey tube, the system airflow is not to exceed 85 LPM.
CAUTION–
CAUTION–
CAUTION–
U s e I n s t r u c t i o n s
Figure 3
U s e I n s t r u c t i o n s
OPERATION
CAUTION–
CAUTION–
CAUTION–
NOTE–
NOTE–
WARNING
• Should the airflow be stopped or interrupted, take off the mask and turn off
the 9200. When the airflow is resumed, turn on the 9200 and allow 20 min-
utes for it to stabilize.
• Do not touch the Heater Plate. The Heater Plate temperature may exceed
65°C (149°F).
• Do not touch the underside of the Humidifier Chamber when it is removed
from the hot Heater Plate. The base will take some time to cool.