The pump:
When the regulation device must be stored, clean water should be pumped through the hose in
Then position the roller arm at 45°, turning it clockwise.
order to rinse it.
These two precautions will facilitate subsequent reactivation of the device.
Keep away from freezing conditions.
The probe:
Extract the Redox probe from the relative probe holder. Replace it in the original bottle filled with tap water. If
necessary, close the probe holder using a plug the size of a 5 euro cent coin.
!
Keep away from freezing conditions, intense heat and direct sunlight.
!
DO NOT INSERT EXCESSIVE QUANTITIES of chemical product above the probe.
As the Redox electrode is comprised of glass elements, handle it with care.
OFA (overdose alarm):
First OFA alarm: the ALARM light flashes; activation after 3 consecutive dosing cycles in which the set value was
not reached; the system continues to guarantee the measurement and dosing functions.
Second OFA alarm: the ALARM light flashes and the Hold light comes on; the dosing pump is blocked. The alarm is
activated after 4 consecutive dosing cycles in which the set level was not reached; to return to normal operation,
press the
: button: the device resets the alarms and returns to normal measuring and dosing mode.
Flow: presence of incoming flow (enslavement by the filter pump).
Alarm
Level
Out Range measure
OFA First Alarm (time >28 min)
OFA Second Alarm (time >40 min)
Flow Rate
Calibration Function
System Error
Default parameters:
Set Point value= 750 mV
•
Redox Range = B
•
Calibration = ON
•
OFA = OFF
•
Flow Rate= ON
•
0000136865
All manuals and user guides at all-guides.com
The alarms
Leds
Alarm led Flashing
Hold led light ON
Alarm led Flashing
Alarm led Flashing
Alarm led Flashing
Hold led light ON
Hold led light ON
Alarm led Flashing
Hold led light Flashing
All leds Flashing
To restore Default parameters run Following steps:
Power off Pool Basic unit
•
Keep Press
•
The unit flashing all leds
•
Press
•
rev. 6.0
Relay
Alarm Relay Close
- Push
Alarm Relay
- Restore Product tank
Alarm Relay Close
- Push
Alarm Relay
- Replace Measure pH
Alarm Relay open
- Push
Alarm Relay Close
- Push
Alarm Relay open
- Restore Flow Rate
Alarm Relay open
- Restore Probe or Buffer
solution and repet
calibration function
Push
Alarm Relay open
initialization
with
Key and Power On
Key to restore Default parameters.
Actions to do
Key to open
Key to open
Key to reset
Key to reset
Key to retry