below:
For 230V, 50Hz - T1A
For 120V, 60Hz - T1.6A
4.
Refit the base and four securing screws.
5.
It is advisable to perform a 'PAT' (Portable
Appliance Test) electrical safety test, or
equivalent, before the unit is put back into
operation.
For a comprehensive list of parts required by service
engineers conducting internal repairs, or to
receive a service manual, please contact the
Service Department at;
cpservice@antylia.com
Tel: +44 (0)1785 810475
Quote both the model and serial number.
Only spare parts supplied or specified by Antylia
Scientific or its agents should be used. Fitting of
non-approved parts may affect the performance and
safety features designed into the instrument. If in
any doubt, please contact the Service Department of
Antylia Scientific or the point of sale.
Technical Specification
Number of samples
Temperature range
Temperature resolution
Temperature accuracy at 20ºC
Temperature accuracy at 300ºC
Display
Sensor
Ramp rates
Dimensions (WxDxH)
Net weight
MP-200D
Two
Ambient to 300°C
1°C
±1.0°C
±2.5°C
Three digit LED
PT100
20°C / minute to plateau
2°C / minute to melt
160 x 220 x 170 mm
1.8 Kg
MP-200D-HR
Two
Ambient to 300°C
0.1ºC
±1.0°C
±2.5°C
Four digit LED
PT100
20°C / minute to plateau
Between 1 and 10ºC/ min
160 x 220 x 170 mm
1.8 Kg
5