Operational Precautions and Limitations
WARNINGS:
CAUTION: Risk of Danger. The illumipro-10 is an electromechanical device
that can cause physical shock or injury to the operator if not used in accordance
with the procedures described in this manual.
LASER RADIATION: Avoid Exposure to Beam. The illumipro-10 contains a
class 3R laser product. The laser will not function when the lid is in the open
position,
however, care should be taken in the handling and use of this instrument.
HOT SURFACE: Keep hands Away from Hot Surfaces. The illumipro-10
contains a heat block producing temperatures between 55 – 65 C during
operation. Care should be taken to avoid direct contact with the heat block.
CAUTION: Laser Radiation.
Service of the unit should be performed only by qualified personnel as optical
exposure to the laser beam may cause injury.
CAUTION: Protect from water. The illumipro-10 does not protect from the
IPX-0
egress of water. Do not expose to water or submerge the instrument.
PRECAUTIONS:
•
The illumipro-10 is an automated instrument that utilizes isothermal Loop-Mediated Amplification
Technology. Care must be taken to avoid contamination of the equipment and workspace by
target and/or amplified nucleic acids. Only qualified personnel should perform molecular testing.
•
The illumipro-10 is used in conjunction with Meridian Bioscience, Inc's illumigene Loop-
Mediated Amplification in vitro diagnostic products.
should be handled in accordance with the instructions provided in specific illumigene product
instructions for use.
•
Selected Language cannot be changed while a run is in process.
•
Time and Date cannot be changed while a run is in process
•
Optical System Verifiers must be stored in the case provided. The optical verifiers must be
protected from light and damage. Optical system verifiers should not be stored in the illumipro-
10.
•
Do not hot swab the optional illumpro-10 keyboard. CAPS Lock will come on at power-up to
indicate keyboard is online.
•
The optional Keyboard must be plugged in while the illumipro-10 is powered-off.
•
When not in use, the illumipro-10 should be stored with the lid closed.
illumipro-10 Operator's Manual: ENGLISH
Copyright© Meridian Bioscience, Inc.
SN11007, REV. 07/2010
All manuals and user guides at all-guides.com
The illumipro-10 contains a laser product.
Samples processed in the illumipro-10