SAFETY PRECAUTIONS

Wait for at least 15-20 minutes after drinking alcohol or
smoking cigarette and before you start the test. Remember
to rinse your mouth with water.
For more precise results, it is advisable to calibrate the device
every 12 months or 300 tests.
The device operates at the temperature of 10 – 40°C, with optimal
precision for tests at room temperature (i.e. ca. 20 – 26°C).
If the test result is lower than 0,10 ‰ or 0,10 % BAC or 0,050
mg / L, it is natural if you did not drink alcohol before. It
happens because in some cases it is possible to measure
other substance than alcohol.
Do not use the device to specify and decide about driving
vehicles. The alcohol level result is for information purposes
only. The manufacturer and retailer waive any legal liability
for the estimated/measured test value.
MAINTENANCE
To keep the device in a good condition, it is advisable to use the
device at least once a month.
The sensor of the device is not ready. It happens
when there is high alcohol or dirt remains from the
last test on the sensor. This being the case, switch
the device off and wait for a couple of minutes
before you take another test.
The device was unable to finish the test. Retake it.
The temperature of the test is too low or too high.
Restore room temperature for the device. Before
you take another test, wait for a couple of minutes
until the sensor inside the device reaches the room
temperature.
Low battery. Use new AA alkaline batteries.