TREATMENT OF THE WHOLE RESPIRATORY TRACT
1. INSTRUCTIONS FOR RF6 BASIC
Prior to use, the nebuliser and the accessories must be
cleaned according to the instructions provided in the
"Cleaning, disinfection and sterilisation" paragraph.
Remove the unit from the carrying bag (13) before use!
1 Plug the power cord (7) into the socket (6) at the rear of
the unit and plug the power cord into a power socket
corresponding to the voltage of the unit.
2 Carefully wash your hands before going ahead with
inhalation therapy.
3 Open the nebuliser by turning the upper part (9c)
anticlockwise.
4 Pour the drug prescribed by the physician into the lower
part (9a). Close the nebuliser by turning the upper part
(9c) clockwise.
5 Connect the accessories as shown in the "Assembly
diagram". The upper cap (9d) provided with your
"RF6 basic
nebuliser makes it possible to deliver the
2
medication in 2 modes.
Standard mode FIg. A1
For an excellent delivery with a minimum waste of
medication.
High speed mode FIg. A2
For a quicker application.
6 Sit comfortably, holding the nebuliser in your hands, place the mouthpiece over your mouth or use
the mask. Start the unit by means of switch (5) and breathe in and out deeply.
7 When the treatment is finished, turn off and unplug the unit.
8 Clean the nebuliser and the accessories according to the instructions provided in the "Cleaning,
disinfection and sterilisation" paragraph.
9 It may happen that after a treatment session visible humidity deposits form in the tube (8): disconnect
the tube from the nebuliser and let it dry using the compressor air flow, as this prevents mould from
growing in the tube.
2. CLEANING, DISINFECTION AND STERILISATION
Before undertaking any cleaning operation, switch off and unplug the unit.
Accessories for aerosoltherapy and Nasal wash
• Cleaning: Wash under running water with mild (non abrasive) washing-up liquid or in the hot cycle
dishwasher, after disassembling all the parts (except for the connecting tube). Clean the nozzle (9b)
after detaching it from the upper part (9c) of the nebuliser and disassembling as shown in FIg. A3.
• Disinfection: All accessories (except for the connecting tube) can be disinfected or sterilized choosing
one of these methods: a) by soaking in a solution of water and vinegar (60% water and 40% vinegar).
Finally, rinse thoroughly with drinking water; b) by boiling in water for 20 min.; this method cannot be
used for masks. Use preferably distilled or demineralised water.
• Sterilisation: Sterilise all parts with cold chemical sterilizers (peracetic acid based, following the
instructions provided by the manufacturer), or (except for tube and masks) in an autoclave at 121°C
(1.2 bar) for 20 minutes, carefully following the instructions provided by the autoclave manufacturer.
• Drying: After washing or disinfecting the accessories, dry them very carefully with a paper tissue or,
better yet, a warm air flow (e.g. from a hair dryer).
Reassemble the nebuliser nozzle (9b) and fit it in the nebuliser as indicated in the "Assembly diagram"
FIg. A.