Before loading the machine and before starting a
programme
- Check that the mobile unit is correctly docked to the water path in
the chamber.
- Are the inserted modules correctly connected to the water supply
of the mobile unit?
Risk of fire due to overheating
When the APLW 511 mobile unit is used without modules, all water
connections are closed. During drying this will cause a build-up of
hot air. The drying units can overheat and catch fire.
The APLW 511 mobile unit must only be used without modules in
programmes that do not include a drying programme block.
Reprocessing programmes
New reprocessing programmes adapted for the mobile unit's loading
templates are available for the APLW 511.
The new reprocessing programmes are available from the manufac-
turer of the cleaning machine.
The factory default setting assigns the standard programmes for the
PLW 8615, PLW 8616 and PLW 8617 cleaning machines to the A 503
mobile unit. The standard programs must be copied and adapted for
the APLW 511 mobile unit.
Modules
Up to 9 modules on 3 levels can be accommodated in the APLW 511
mobile unit. Among other things, the possible number of modules de-
pends on the height of the load items on the bottom and middle
levels of the mobile unit.
Examples:
For reprocessing long pipettes, 3 modules are placed on the bottom
level.
For reprocessing small items of laboratory glassware, 3 modules can
be placed on each level.
For reprocessing phials, one A 605 module can be placed on each
level. The three A 605 modules must be arranged diagonally, not one
on top of the other. A 620, A 621 or A 622 modules can be added in
the available space on each level.
en - Areas of application
33