EMC DECLARATION
The ME EQUIPMENT (NASAL ASPIRATOR) or ME SYSTEM is suitable for home
healthcare and clinic environments.
WARNING: Do Not use this equipment near active HF surgical equipment or the RF
shielded room of an EM magnetic resonance imaging system, where the intensity of
EM disturbances is high.
WARNING: Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that
they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30cm
(12 inches) to any part of the equipment NC003, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
1.
All necessary instructions of maintaining BASIC SAFETY and ESSENTIAL
PERFORMANCE regarding electromagnetic disturbances for the expected
service life.
Portable and mobile RF communication equipment may affect the performance
of the NC003, avoid strong electromagnetic interference when in use, such as
near mobile phones, microwave ovens, etc.
2.
Guidance and manufacturer's declaration - electromagnetic emmissions and
immunity.
GUIDANCE AND MANUFACTURER'S DECLARATION -
ELECTROMAGNETIC EMMISSIONS AND IMMUNITY
Table 1
Guidance and manufacturer's declaration - electromagnetic emmissions
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable