Release the Power
button when the desired temperature scale is shown.
There will be a short beep to confirm the new setting, then the thermometer is turned off
automatically.

Care and cleaning

The probe tip is the most delicate part of the thermometer.
It must be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter, clean the
probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened
with alcohol. After the alcohol has completely dried out, you can put a new
lens filter on and take a temperature measurement.
If the probe tip is damaged, please contact Consumer Relations.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and lens filters
in a dry location free from dust and contamination and away from direct sunlight.
Additional lens filters (LF 40) are available at most stores carrying Braun ThermoScan.

Replacing the batteries

The thermometer is supplied with two 1.5 V
batteries. For best performance, we recommend Duracell
batteries.
1
Insert new batteries when the battery symbol appears on the display.
2
Open the battery compartment. Remove the batteries and replace with
new batteries, making sure the poles are in the right direction.
3
Snap battery door into place.
Only discard empty batteries. To protect the environment, dispose of
empty batteries at your retail store or at appropriate collection sites
according to national or local regulations.
Calibration
This device was designed and manufactured for a long service life, however it is generally
recommended to have the device inspected once a year to ensure correct function and accuracy.
Please contact the authorized service centre located in your country.
Note: The calibration checking is not a free service. Please contact the Authorized Service Centre to
get a quotation before you send out the product.
Manufacturing date is given by the LOT number located in the battery compartment.
The first three (3) digits after LOT represent the Julian date that the product was manufactured and
the next two (2) digits represent the last two numbers of the calendar year the product was
manufactured. The last identifiers are the letters that represent the manufacturer.
An example: LOT 11614k, this product has been manufactured on the 116th day of the year 2014.
8
All manuals and user guides at all-guides.com
type AA (LR 06)
alkaline
®
Errors and troubleshooting

Error Message

Situation
No lens filter is attached.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low
System error – self-check display
flashes continuously and will not be
followed by the ready beep and the
ready symbol.
If error persists,
If error still persists,
Battery is low, but thermometer will
still operate correctly.
Battery is too low to take correct
temperature measurement.
Solution
Attach new, clean lens filter.
Take care that the positioning of the
probe is correct and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new
measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10
and 40 °C or 50 and 104 °F .
Make sure the probe tip and lens are
clean and a new, clean lens filter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
... reset the thermometer by removing
the batteries and putting them back in.
... please contact the Service Centre.
Insert new batteries.
Insert new batteries.
9