If it takes longer than 20 seconds to obtain a blood
sample and to touch the test strip to the blood drop,
repeat Step 3.
Step 4:
The meter will count down, showing a flashing
dash. After 7 seconds, a beep sounds and the test result
and date/time are shown.
Results will be shown in units of mg/dL.
Record result in your logbook.
Step 5:
When the test is done, pull out test
strip. Dispose of used test strip. The meter
shuts off automatically when you remove
the test strip.
Remove lancet from lancing device. Dispose
used lancet according to local guidelines.