The infrared thermometer has been tested and conforms to the standard
ASTM E1965-98. ASTM laboratory accuracy requirements in the display
range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is
±0.4°F (±0.2°C). Note that for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is ±0.2°F (±0.1°C).
Full responsibility for the conformance of this product to the standard is
assumed by Shenzhen Jumper Medical Equipment Co., Ltd.
(Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan,
Shenzhen,Guangdong,China)

Security Class

Type of protection against electric shock: internally powered equipment.
Degree of protection against electric shock:
■ Degree of protection against ingress of water:IP22
■ Safety degree of using in flammable anesthetic gas blending with air,
oxygen or nitrous oxide: Non-AP/APG
■ No application parts of the thermometer prevents defibrillation charge
effect.
■ No application parts of the thermometer output signal.
■ The thermometer is impermanent installed device.

Storage and Transportation

The thermometer can be transported using general transportation tools.
Severe vibration, shock, or rain must be avoided during transportation.
Atmospheric pressure: 86–106 kPa
Type BF applied part.
28