NOTE:
If using only remote START/STOP, PRIME and/or CW/CCW, the MODE control can be set to any of the
three positions.
5. To adjust the voltage or current scaling for other than zero to full scale, Press and hold the MODE key while
pressing the FLOW key. This display will show "LO" for 2 seconds and then a flow rate. Use the INC/DEC ( ,
arrow keys to set the minimum control level. Press the FLOW key again. The display will show "HI" for 2 seconds
and then a flow rate. Use the INC/DEC ( ,
exit. The same scaling will be used for both input and output levels.
NOTE:
Tubing size should be selected before adjusting the voltage or current scaling.
To provide sufficient hysteresis between minimum speed and off, a 0.2 mA and 0.1 V offset is added to the current
and voltage inputs and outputs as shown in the following table. This offset is only added when the input/output
scaling is set with a minimum speed of zero (default)
Speed (rpm)
4-20 mA In/Out
0
4-4.2 mA
1
4.2263 mA
600
20 mA
PRIME
CW/CCW
Figure 2
0-20 mA In/Out
0-0.2 mA
0.233 mA
20 mA
RED/YELLOW
BLUE
GREEN
START/STOP
YELLOW
WHITE
INPUT
0-20mA; 4-20mA
ORANGE
INPUT
0-10V
BLACK
BROWN
OUTPUT
0-20mA; 4-20mA
VIOLET
OUTPUT
0-10V
RED
) arrow keys to set the maximum control level. Press any other key to
0-10 V In/Out
0-0.1 V
0.1165 V
10V
NOTE: Colors are those of
Remote Cable 77300-32.
14
6
4
GREY
10
TAN
8
12
PINK
3
11
RED/GREEN
5
13
RED/BLACK
1
15
16
7
N.C.
9
17
N.C.
2
18
N.C.
07575-00
TACH OUTPUT
MOTOR RUNNING
N.O. CONTACT
MOTOR RUNNING
N.C. CONTACT
)
7