10 Repairs
The sensor may only be repaired at the factory. Please return in its original
packaging. Prior to return, please carry out instructions as described in
11 "Decommissioning".
11 Decommissioning
Decommissioning sensor: observe all safety guidelines as given in section 5
"Assembly"!
S Disconnect sensor from power supply (see 6 "Installation")
S Depressurise the in-line probe housing
S Loosen the locking screw
S Slowly remove the sensor from the in-line probe housing
S Unscrew the membrane cap over a wash basin or similar and empty
S Rinse the membrane and electrodes with clean water and allow to dry (do
not allow dust to settle on electrodes or membrane)
S Screw on a fresh membrane cap loosely to protect the electrodes
S Fit the membrane cap cover to protect the membrane cap
12 Disposal
Electrolyte
The electrolyte can be disposed of with in-house waste.
Sensor
ProMinent
®
Repairs / Decommissioning / Disposal
IMPORTANT
Electronic waste is classified as special waste!
Observe currently applicable local directives!
31