Micro 800 pH mV Operating Instructions
Getting Started/Connections
After installing (2) AA batteries and/or connecting
the optional 110/220 VAC power supply, connect
the desired sensors to the corresponding ports.
BNC: pH, mV/ORP
or Ion Selective
ATC/Con: 6-Pin ATC
or ATC/Conductivity
A/C Power
Adapter
12mm and 16mm probes can utilise the probe
holder to attach one or more sensors to a beaker
and to the instrument as needed. The stand can be
extended as shown above or used for wall-mounting.
Keypad Functions
Press once to power ON in the mode that was
previously used. Press again to turn backlight on for
one minute or off. Hold for 3 seconds to power OFF.
Toggle between measurement and calibration modes.
In SETUP mode, BACK serves to return to the previous
menu option or setting.
Confirm calibration values in CAL mode.
Confirm selections in SETUP mode.
Freeze or release the measured reading.
Customise instrument settings and preferences.
(See also Setup Programs).
Toggle between available measurement types.
Save measurement into memory. Increase value
or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value
or scroll down in SETUP or manual calibration.
Send output data to printer or computer.
2
Setup Programs
To access the settings below, press SETUP. Up/down
arrows will display the available options. Press
ENTER to accept the desired setting, or BACK
to return to the previous option and/or exit.
Configuration Options
• Ready indicator ON/OFF/
or Automatic HOLD when stable.
• Choose ºCelsius or ºFahrenheit.
pH Buffer Options
• Select the desired pH Buffer Calibration Group:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or
NIST (1.68, 4.01, 6.86, 9.18, 12.45) or
DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or
MAN (manual adjustment of any custom
pH values that are ≥1 pH unit apart.
• Select number of calibration points.
Output (USB or
RS232 - PC or Printer
Select Calibration Due Reminder
Communication)
• Set number of days from 0-60 for desired parameter.
View Calibration Data
• Press ENTER to view each point that is calibrated.
View Electrode Data
• Press ENTER to view mV Offset and
Slope % of the measured reading.
System Settings
• Data Logging:
MANUAL upon key press only TIMED interval.
Choose (SEC/MIN/HOUR) interval.
• Automatic shut off after 10 minutes.
Choose ON or OFF.
• Clock Settings:
Date: Choose USA (MM/DD/YYYY)
or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR).
If 12HR, choose AM or PM.
• Set Printer Type:
CSV (Comma Separated Values) -
best format for computer.
Printer (Text) - best format for printer.
Choose Manual (MAN) upon
key press or TIMED interval.
If timed, choose (SEC/MIN/HOUR).