Atmospheric Pressure 700hPa - 1060hPa
Maximum Operating Pressure 60hPa
46
Patient category
Neonatal
8. Directions for Use
(1) Open the product package by tearing off side opening. DO NOT use scissors or sharp instruments to open the package.
(2) Before set-up, check tubing and connectors for damage and that all connections are secure.
6. Product Structure
Neonatal Single Heated Breathing Circuit with Water Trap
(3) When mounting a humidifier adjacent to a patient ensure that the humidifier is always positioned lower than the patient.
Neonatal Single Heated Breathing Circuit with Water Trap
(4) Connect the Ventilator to Humidifier Chamber Limb to the ventilator and the Heated Humidifier chamber.
(5) Connect the Inspiratory Limb (Blue) to the Heated Humidifer chamber.
(6) When used with an incubator the incubator limbmust be used to allow the temperature probe to sit outside the incubator.
Incubator limb
When used without an incubator the incubator limb must be removed from the circuit.
(7) Connect the Expiratory Limb (White) to the Ventilator.
15F
(8) Insert both the chamber and patient probes ensuring that both probes are pushed home.
Swivel wye
(9) Connect the heating wires of the Heated Humidifier to the circuit.
(10) Connect the pressure monitoring line to the ventilator.
(11) Verify that all connections are properly made.
15M
(12) Perform a pressure and leak test on the breathing circuit and check for occlusions before connecting to a patient.
(13) Make sure all ports like sampling, pressure measuring and temperature are closed when they are not in use.
Ventilator to humidi er chamber limb
(14) Patients should be monitored constantly whenever they are using the product.
(15) Check the product periodically; it must always be replaced when it becomes unfit for its intended use.
(16) When the product is not connected to the patient, protect the patient side of the product to prevent the intrusion of foreign bodie
(17) After using the circuit, remove it from the humidification system and discard in accordance with applicable hospital requirement
or local laws and regulations.
Ventilator
Ventilator to humidi er
chamber limb
Humidi er
9. Warnings
(1) Do not stretch or milk the tubing. Do not crush the tube.
(2) Do not cover the breathing circuit with materials such as blankets, towels or bed linen.
(3) DO NOT STERILISE. Soaking, rinsing or steam sterilization of the product may leave harmful residues on it and lower its performance.
(4) Avoid contact with chemicals, cleaning agents or hand sanitizers.
(5) In the unlikely event in certain environmental condition condensation (rain out) may form. In these conditions turn off the flow and
disconnect the circuit. Drain off any excess fluid or change the circuit. Reconnect as necessary.
AW6527(1)
(6) This product is intended to be used for a maximum of 7 days.
Leakage limit
≤ 30ml/min@(60±3)cmH
O
≤ 0.74cmH
2
Pressure monitoring line
Water trap
22F
Pressure monitoring
line
Expiratory limb
Inspiratory limb
Chamber probe
Heated wire cable
This product is intended to be used for a maximum of 7 days.
Instruction For Use - Neonatal Single Heated Breathing Circuit with W
Flow resistance limit
Compliance limit
O/l/min@2.5 l/min ≤ 1.5ml/cmH
2
4-6M
Inspiratory limb
Expiratory limb
15F - 9.5F
7.5M - 10M
6M - 15M/12F
22M/15F
22M/15F
Straight adaptor
Restrictor
Incubator limb
Patient probe
Incubator
Circuit length
O@(60±3)cmH
O
1.6m
2
2
22F
15M
Universal connector pack
4M - 3M/4M
12M - 12F
22M/15F - 15M
11F - 15M/12F
8M - 15M/12F
Patient temperature probe
Patient temperature probe
15F - 9.5F
15F - 9.5F
02.02.2021