6.9 Calibrating pH-2004

The pH-2004 need the following standards for calibration: (See 1.2)
T ª
pH 1
20º C
1. 6 7 5
25º C
1. 6 7 9
30º C
1. 6 8 3
Calibrate in one or two points using one or two of this buffers.
Choose this two point by considering:
One point is always pH=6.85
If you are going to measure an acid solution choose pH=4.00
If you are going to measure an alkali solution choose pH=9.18
Calibrating pH-2004:
Clean the electrod e with water.
From the measuring screen, press «MODE/MEAS»
Press «
Immerse the electrode on the buffer solution pH=6.86 at 25ºC
Wait for stable pH reading.
Reading can be switched to mV by key "pH/mV".
Press «ENTER»
The display shows the recognized buffer value.
This value can not be changed.
After 5s. The unit jumps, automatically to STD2 screen.
Press «ENTER»
If calibration its only in 1 point, that's all, press «MODE/MEAS»
To continue for a two points calibration (strongly suggested) press
«ENTER» on «STD2» status.
Clean, rinse, the electrode with distilled water.
Immerse the electrode on the buffer solution pH=4.00 or 9.18 a 25ºC
Wait for stable pH reading.
Reading can be switched to mV by key "pH/mV".
Press «ENTER»
The display shows the recognized buffer value.
This value can not be changed.
After 5s. ..
The calculated SLOPE value is shown
This value can be used to evaluate the electrode state:
Slop e:
9 8 - 1 05 %
94 -9 7 %
9 0 - 9 3 %
Nota: Calibrate pH-2004 when:
First time use
Daily.
After change the electrod e
After measuring strong acids (pH<2)
After measuring strong alkalis (pH>12)
After measure very dense solutions that can obstruct the membrane.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
(Sujetas a modificaciones sin previo aviso)
pH 4
pH 7
4. 0 0 2
6. 8 8 1
4. 0 0 8
6. 8 6 5
4. 0 1 5
6. 8 5 3
» to get «STD» screen validate with «ENTER».
electrod e state:
Electrode is good.
Clean the electrode with water and calibrate again.
Bat state /repair or change the electrode. See 6.15
pH 9
pH 12
9. 2 2 5
1 2 . 6 2 7
9. 1 8 0
1 2 . 4 5 4
9. 1 3 9
1 2 . 2 8 9
Tel 34 937 700 877 Fax 34 937 702 362
Pág..: 17