Measuring the concentration
Measuring using reagent tests
After approx.
2 s:
12
– Insert the AutoSelector into the round cell
shaft.
Align the line mark to the notch of the pho-
tometer.
The photometer reads the barcode and auto-
matically selects the relevant method.
– Insert the rectangular cell (vertically, to the
right of the slot). The measurement starts
automatically.
Observe the position mark at the cell shaft.
If the select method menu is displayed,
insert the corresponding AutoSelector
into the round cell shaft.
If the "cell type" query appears, select
the relevant rectangular cell (10 mm,
20 mm, 50 mm) with
confirm with
The measured value appears on the display.
or
,
.