Spectroquant
NOVA 60A - Operating Manual
®
Maintenance, cleaning, disposal
Cleaning - Actions to take if a cell is broken
Do not rotate the photometer to
pour out the liquid!
The photometer has a draining mechanism
under the cell shaft that, when operated cor-
rectly, prevents any liquid coming into contact
with electronic components.
– Switch off the photometer (close the cover)
and disconnect it from the line power
– Let the liquid drain off
– Carefully remove any pieces of glass, e. g.
using tweezers
– Carefully clean the cell shaft with a damp,
lint-free cloth
– Let the cell shaft dry.
Disposal
Packing
The measuring instrument is sent out in a pro-
tective transport packing.
We recommend: Keep the packing material in
case you have to send the measuring instru-
ment back for service.
The original packing prevents the measuring
instrument from being damaged during trans-
port.
Measuring instrument
Dispose of the measuring instrument as elec-
tronic waste at an appropriate collection point.
It is illegal to dispose of it in household refuse.
Within the European Union, the batteries are
removed at a specialized treatment center at
the instrument's end of life. The instruments
are taken to one of those specialized treatment
centers via the recycling system set up for this
purpose.
18
After it is dry, check the photometer:
– Perform a photometer monitoring (see section
PHOTOMETER MONITORING (AQA1), in the
FUNCTIONAL DESCRIPTION document [on
CD-ROM]).
Release 06/2019