19. GENERAL SPECIFICATION

Waveform
Amplitude
FACE 1 over 500 Ohm load
FACE 2 over 500 Ohm load
Other programmes over 1
kOhm load
Output plug
Channels
Batteries
Weight
Dimensions
Safety Classification
Environmental specifications:
Operating:
Storage:
TYPE BF
EQUIPMENT
IP22
Note: The electrical specifications are nominal and subject to variation from
the listed values due to normal production tolerances of at least 5%.
Note: At least 30min required for the device to warm / cool from the minimum
/ maximum storage temperature between uses until it is ready for intended
use.
Asymmetrical Bi-Phasic
40 mA zero to peak +ve in 40 steps.
300 µA in phase 1, 250 µA in phase 2 and 80 µA in
phase 3.
90 mA zero to peak +ve in 90 steps +/- 10%
Modified constant charge control
Fully shielded: touch proof
Dual channel
2 x AA alkaline (two AA batteries)
75 g without batteries
120 x 60 x 20 mm
Internal power source.
Designed for continuous use.
No special moisture protection.
Temperature range: 5 to 40⁰C
Humidity: 15 to 93% RH non-condensing
Atmospheric pressure: 700 hPa to 1060 hPa
Temperature range: -25 to +70⁰C
Humidity: Up to 93% RH non-condensing
Atmospheric pressure: 700 hPa to 1060 hPa
Equipment providing a degree of protection against
electric shock, with isolated applied part.
This symbol on the unit means "Refer to Instructions for
use".
The unit is not water resistant, and should be protected
from liquids.
Complies with EU WEEE regulations
33