B4 Troubleshooting - KaVo eXam 0.805.0333 Manual Del Usuario

Sistema de radiología intraoral digital
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

B. Installation and setup

B4 Troubleshooting

PROBLEM
The unit does not come on. The unit's power on / off
status light and display are off.
CAUSE / SOLUTION
The main power supply to the unit is off or the unit is
not switched on.
i.
If the LED on the PSU is not on it is not receiv-
ing power from the mains. Switch the mains
power on.
ii. If the LED on the PSU is on switch the unit on.
iii. If status light and display still do not come on
check the unit's membrane control panel and
cabling. Replace if faulty.
PROBLEM
Animation displaying either Unit connection or
Dental imaging software (fl ashing yellow) appear
on the unit display and the unit will not work.
CAUSE / SOLUTION
Defective RJ45 cable. Replace.
Imaging plate unit
B - 15