LED
3.3 Overview interfaces
1
Connector
1
Socket for AC/DC
adapter
2
USB-B port (to
host)
3
RS232 interface
8
Design and Function
State
One LED lit up
Two LEDs lit up
Three LEDs lit up
All four LEDs lit up
First LED flashes
Fourth LED flashes
All Four LEDs flashes
2
Function
To connect the power supply. Use only the +12 V DC standard AC/DC
adapter which is delivered with the accessory.
To connect a balance, a personal computer or a hub provided with an USB
port.
The connection for RS232 is a 9-pin D-sub miniature connector, pin
assignment as for computers. For the attachment of a computer, a so-called
laplink or null modem cable is needed.
Interface parameters:
9600 baud, no parity, 8 data bits, 1 stop bit
Software handshake Xon/Xoff
Line ending <CR><LF>
Meaning
Lowest feed rate
Lower intermediate feed rate, default after
switching on
Upper intermediate feed rate
Highest feed rate
Short circuit at Start / Stop / Discharge
button
Short circuit at Feed rate / Test button
Instrument is in discharging mode
3
Automatic Feeder