Devices with an earth terminal must always be grounded.
If an earth connection (green-yellow laboratory socket) is available, it must always be
connected to protective earth. Protective earth must always be connected first (before voltage),
and must always be disconnected last (after voltage).
Some devices have a high leakage current. These devices must be additionally grounded
with a protective earth conductor.
The device is not equipped with an integrated fuse unless specified otherwise in the technical data.
Always pull on the plug when disconnecting connecting cables; never pull the cable.
Risk of death in case of interrupted protective earth conductor!
The protective earth conductor (yellow/green) must never be interrupted,
either inside or outside the device.
The insulation of the protective earth conductor must never be damaged or removed.
National regulations for operating electrical systems and equipment must be observed in commercial
facilities.
In schools and training facilities, the operation of power supply units must be responsibly monitored by
trained personnel.
Caution!
The capacitors in the device may still carry a charge, even after it has been disconnected from all
voltage sources.
When replacing fuses, only use specified fuses with the correct current rating.
Never switch the device on immediately after it has been moved from a cold room to a warm room.
Under unfavourable conditions, the condensate which forms as a result may damage the device. Leave
the device switched off until it has reached room temperature.
Risk of death from connecting power supply units in series!
Contact voltages > 25 V DC are not permissible.
Coming into contact with voltages of greater than 120 V DC may be fatal.
Do not connect voltage sources in series.
Device ventilation slots must not be covered! The devices must be placed on hard, flame-resistant
surfaces so that air can enter them unhindered. The devices are cooled primarily by means of
convection.
Set the devices up so that activation of switches and disconnectors is not made difficult.
When working on a live circuit, use only suitable tools.