Operation
Measured value
Sign
Unreliability
indicator
[ON/OFF]
The measured value is stable as soon as the unit, e.g. "g", appears. The unreliability indicator lights
up while the scale is recalculating the reference weight, for example, during counting, or when it is
verified and in the warm-up phase.
Application example: Adding up components (application net total)
If the balance is not set to English: Enter the configuration menu, switch the balance off and on with
[ON/OFF] and, as soon as all segments of the display light up, press and hold [MODE] until
"SPRACHE DEUTSCH/LANGUE FRANÇAISE" appears. Press [MODE]; "SPRACHE DEUTSCH/LANGUE
FRANÇAISE" starts flashing. Press [MODE] repeatedly until "LANGUAGE ENGLISH" appears and
confirm with [PRINT]; "LANGUAGE ENGLISH" stops flashing. Press [PRINT]; the balance finishes
starting up.
To enter the application menu, keep [MODE] pressed until "SET APP: OFF" appears. Press [MODE];
"OFF" starts flashing. Press [Down] repeatedly until "NET-TOTAL" appears and confirm with
[PRINT]. Keep [MODE] pressed until "SETUP APPLICATION" appears. Keep [MODE] pressed until
"AUTO-START ON" appears. For manual value transfer press [MODE]; "ON" starts flashing, press
[MODE]; "OFF" appears, confirm with [PRINT]; "OFF" stops flashing. Press [PRINT]; the balance
exits the application menu. Press [MODE]; the balance starts the application Net total and displays
0g as the measured value and "STO 0
Place the first component, e.g. 50g, and press [REF]; the balance records the value, tares itself and
displays the capacity of the weighing range analogically in the info line. Place the second
component, e.g. 100g, and press [REF], and finally place the third component, e.g. 20g, and press
[REF].
While adding up, the current total weight can be viewed at any time by pressing [MODE]. The
balance then switches to weighing mode. To return to the application, press [MODE] again. To
switch from the analogue display to the application commands in the info line, press [MODE] twice;
the balance switches to the weighing mode and back to the application. It then displays e.g.
"STO 3
RES INF".
Press and hold [REF] until "INF" starts flashing; the balance displays in the info line all information
on the weights just added up. Use [REF] to scroll through the entire stack of info lines: Total,
remaining capacity and all individual weights, e.g. 50g, 100g and 20g. Press [MODE] to exit the
information query; the balance switches back to the running application with analogue display and is
ready for another weighing-in.
To restart, press and hold [REF] until "RES" starts flashing; the balance remains in the application
with analogue display and resets the counter to 0.
To quit the application, press [MODE]; the balance switches back to weighing mode.
Auxiliary display for check and reference weighing
[MODE]
[0/T]
RES INF" in the info line.
[REF]
[PRINT]
Measuring unit
Net indicator
Info line