E N G L I S H
WHAT IS BOTTOMING OUT?
Bottoming out may occur on a Fusion Cushion if you displace
the fluid or air underneath your pelvic bones or buttocks, which
would leave you sitting on the foam base. It sometimes occurs
on very thin individuals, people using recliner wheelchairs,
people who have lost weight or have other changes in body size
or shape, or people who slouch when sitting.
When bottoming out occurs, increased pressure is placed
onto the ischials and coccyx increasing the risk for skin
breakdown. Immediately discontinue use of the pad. See
your healthcare professional.
HOW TO CHECK FOR BOTTOMING OUT
JAY Flow Fluid Pad
To check for bottoming out, sit on the cushion, with the cover
on and the zipper opened, for seven to ten minutes. Transfer up
and off the cushion (or have someone help you transfer), trying
not to disturb the fluid underneath you. Fold the back portion of
the cover forward over the front, exposing the fluid pad. Push
down in the depressions on the pad where your ischials (seat
bones) and coccyx (tailbone) were. You should have to push
through at lest 1/2" (1.3 cm) of fluid before you feel the firm
cushion base below.
1/2" (1.3 cm) is approximately this deep.
JAY FUSION CUSHION WARRANTY
Each JAY cushion is carefully inspected and tested to provide
peak performance. Every JAY cushion is guaranteed to be free
from defects in materials and workmanship for a period of 24
months from the date of purchase, provided normal use. Should
a defect in materials or workmanship occur within 24 months
from the original date of purchase, Sunrise Medical will, at its
option, repair or replace it without charge. This warranty does
not apply to punc-tures, tears or burns, nor to the cushion's
removable cover.
The removable cover is guaranteed to be free from defects in
materials and workmanship for a period of six (6) months from
the date of purchase, provided normal use. Should a defect in
materials or workmanship occur within six (6) months from the
original date of purchase, Sunrise Medical will, at its option,
repair or replace it without charge.
Claims and repairs should be processed through the nearest
authorized supplier. Except for express warranties made herein,
all other warranties, including implied warranties of
merchantability and warranties of fitness for a particular
purpose are excluded.
There is no implied warranty beyond what is contained herein.
Remedies for breach of express warranties herein are limited to
repair or replacement of the goods. In no event shall damages
for breach of any warranty include any consequential damages
or exceed the cost of non-conforming goods sold.
©2020 Sunrise Medical JAY Fusion Cushion Owners Manual AU Rev. D
WARNING
If the cushion is properly positioned, the footrests are properly
adjusted, and there is not at least the minimum 1/2" (1.3 cm)
of fluid, the cushion is bottoming out and should not be used. If
you are bottoming out, discontinue use of the cushion and see
your cli-nician. Usually bottoming out is easily solved by
ordering additional fluid in the form of an Overfill. Call your
local authorized supplier to see if this is appropriate for you.
Air Cell Bladder
Reference the Air Insert Pad Owner's Manual Supplement for
instructions on preventing and checking for bottom-ing out.
WHEN TO CHECK
With JAY Flow Fluid, check every month when you clean it.
With Air Cell Bladder, check daily.
JAY Flow Fluid Pad
Consult your healthcare professional for proper fitting and use, and
read the following statements carefully:
Fluid Pad Policy
If a Clinician, Certified Rehab Professional or Authorized Sunrise
Medical Supplier has determined that the fluid volume is
inappropri-ate for the original purchaser, Sunrise Medical, during
the warranty life of the product, will replace the fluid pad free of
charge. Requests must be submitted by an authorized Sunrise
Medical supplier.
• The fluid in this cushion may lose volume over time.
• Volume loss in this cushion may result in bottoming out
the cushion.
• Monthly, during routine cleaning and inspection, check
this cushion for bottoming out (see instructions under
"checking for bottoming out" in the warranty use and
care guide) and any fluid inconsistencies, If the fluid is
firmer in one area sim-ply knead it back to its original
consistency.
• If bottoming out occurs discontinue use of this cushion
and contact your healthcare professional. If further
assistance is necessary, contact Sunrise Medical
Customer Service at 800-333-4000.
Air Cell Bladder
Reference the Air Insert Pad Owner's Manual
Supplement for additional use and care guidelines.